INSTRUMENT/METHOD

INSTRUMENT : gas chromatograph (GC) with capilary column IDENTIFIER : IO481997IN038

Project : MAST III/MTPII/MATER

 

Laboratory :

Prof. Dr. W. Roether


Address :

Universität Bremen, FB 1, Institut fuer Umweltphysik

P.O. Box 330 440

28334 Bremen, Germany

Tel : 49 (0)421 218 3511

Fax : 49 (0)421 218 7018

Email : wroether@physik.uni-bremen.de


Contact Scientist :
Dr. Birgit Klein, adress as above, email: bklein@physik.uni-bremen.de

INSTRUMENT NAME : gas chromatograph (GC) with capilary column

MANUFACTURER : CG : Hewlett Packard 5890, main column: J&W Scientific, Trap filling: Alltech, Valves: Valco Instruments

MANUFACTURING DATE :1992

SERIAL NUMBER : GC Hewlett Packard 5890

DESCRIPTION :
The gaschromatographic system at Bremen has been custom designed to allow high precision measurements of four CFC species (CFC-11, CFC-12, CFC-113 and CCl4). The combination of a capillary column with a low-volume trap yields high resolution in the chromatograms and full seperation of peaks. The system is fully automated and allows eight water samples and associated calibration measurements to be carried out un-attented. Water samples are transferred into the system by means of flow-through containers, incorporating glass ampoules. Generaly, measurements will be carried out immeadiately after sampling at sea, but alternatively the ampoules can be flame-sealed, allowing sample storage for later analysis ashore. Standard gas is used for calibration and all concentrations are reported on the SIO93 scale. The overall precision for surface waters amounts to ± 0.6% (CFC-11, CFC-12) ± 1.5% (CFC-113) and ± 2.4% (CCl4). Detection limits are 0.002, 0.002, 0.008 and 0.003 pmol/kg for CFC-11, CFC-12, CFC-113 and CCl4.

TECHNICAL CHARACTERISTICS :
The gas chromatograph is a Hewlett Packard 5890, the main column GS GASPRO (ID 0.32 mm, 30 m long) is manufactored by J&W Scientific and the trap column (1mm ID, 10 cm long) is filled with Unibeads 2S (Alltech).

COMMENTS :

INSTRUMENT TYPE :Laboratory equipement X

OTHER ATTACHED EQUIPEMENT :

DATE (of updating this form) : 18.10.1999

MEASURED PARAMETER 1

CALIBRATION DATE

CRUISES

COMMENTS

CFC-11, CFC-12, CFC-113

Calibration curve every 3 day, standards every 3 hours during measurement

Urania mai2-97

 
       

BIBLIOGRAPHICAL REFERENCES :
Bulsiewicz, K. , H. Rose, O. Klatt, A. Putzka and W. Roether, A capillary-column chromatographic system for efficient chlorofluorocarbon measurement in ocean waters, J. Geophys. Res., Vol. 103, C8, 15959-15970, 1998.

Last update 15/12/1999